文件编号NR-TP-0715-008 版本1.0
- 83 -
HR
Measuring range
3/5-lead monitoring
Adult: 15bpm~300bpm
Pediatric: 15bpm~350bpm
Resolution
1bpm
Heart rate measurement
error
±1bpm or ±1%, whichever is greater
TEMP
Standards compliant
EN12470-4, ISO 80601-2-56
Measurement method
Thermistor
Measuring range
5~50°C (41~122°F)
Resolution
0.1°C
Measurement accuracy
±0.1°C
Number of channels
Two
Alarm specifications
Range
Step
T1/T2 upper limit
0.1°C~50.0°C
0.1°C
0.1°F
T1/T2 lower limit
C~49.9°C
TD upper limit
0~50°C
RESP
Measurement method
Thoracic electrical bioimpedance method
Measuring lead
Lead I, II
Wave gain
×0.25, ×0.5, ×1, ×2
Respiratory impedance
range
0.5-
Baseline impedance
500-4000Ω
Gain
10 grades
Scan speed
6.25mm/s, 12.5 mm/s, 25mm/s
A.3.2 NIBP Specifications
Standards compliant
EN 60601-2-30/IEC 60601-2-30, EN 1060-1, EN 1060-3, EN1060-4
EN/IEC 60601-1.
Measurement method
Automatic oscillometric method
Operating mode
Manual, automatic, continuous
Useful life
100, 000 times
Measurement interval
1/2/3/4/5/10/15/30/60/90/120/180/240/480min