EasyManuals Logo

Northern Meditec Taurus E User Manual

Northern Meditec Taurus E
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
文件编号NR-TP-0715-008 版本1.0
- 84 -
in automatic mode
Typical measurement
time
20~40s
Normal mode
measuring range
(mmHg)
Adult
Pediatric
Systolic blood
pressure
40-270
40-230
Mean blood
pressure
20-230
20-165
Diastolic blood
pressure
10-210
10-145
Measurement accuracy
Maximum average error: ±5mmHg
Maximum standard deviation: 8mmHg
Resolution
1mmHg
Initial inflation pressure
Default
Pressure setting range
Adult
150mmHg
80~240mmHg
Pediatric
100mmHg
80~200mmHg
Overpressure
protection point
(software)
Adult: 300mmHg
Pediatric: 240mmHg
Overpressure
protection point
(hardware)
Adult: 320~330mmHg
Pediatric: 265~275mmHg
Pressure accuracy
±3mmHg
Electrical characteristics
Supply voltage
10V~14V DC
Maximum power
consumption
3.6w
Quiescent current
50mA
Maximum current
during measurement
180mA
Maximum current
during inflation
300mA
A.3.3 SpO2 Specifications
Standards compliant
ISO 80601-2-61
Display range
0%~100%
SpO2 display
resolution
1%
SpO2 checking
accuracy
2% (70%~100%); not define when lower than 70%

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Northern Meditec Taurus E and is the answer not in the manual?

Northern Meditec Taurus E Specifications

General IconGeneral
BrandNorthern Meditec
ModelTaurus E
CategoryMonitor
LanguageEnglish