A.3 Measurement Specifications
A.3.1 ECG/TEMP/RESP Specifications
EN 60601-2-27/IEC 60601-2-27, GB 9706.25, IEC60601-2-25
I, II, III, aVR, aVL, aVF, V
2.5mm/mV (×0.25), 5mm/mV (×0.5), 10mm/mV (×1.0), 20mm/mV
(×2.0)
6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
50/60Hz notch filter can be set to on or off
Differential input
impedance
Electrode polarization
voltage range
<3s after defibrillation (in monitor and surgery mode)
1mV (peak - peak), accuracy ±3%
For PACE pulses that meet the criteria below, PAEC will be marked on
the screen:
Detection range: ±4mv ~ ±700mv
Pulse width: 0.2ms ~ 2.0ms
Interval of HR
refreshing
Calculate once every second
Time from 80bpm to 120bpm: ≤ 10sec
Time from 80bpm to 40bpm: ≤ 10sec
For T-wave with 100ms QRS wave, 350ms QT period, 180ms duration
and 1.2mV amplitude, the HR calculation won’t be affected