Chapter15 Maintenance
Warning
If the hospitals or institutions using this instrument can’t implement a satisfactory
maintenance schedule, it will result in device failure and may endanger human
health.
15.1 Checking
Check the following basic items before using the monitor:
Check for any mechanical damage;
Check all exposed wires, insertions and accessories;
Check all instrument functions that may be used for patient monitoring and ensure that
the instrument is in good working condition.
If the instrument function has any sign of damage, do not use this monitor for any patient
monitoring. Please contact the hospital's professional maintenance personnel or our customer
service personnel.
Every 6-12 months or after each repair, a comprehensive examination must be performed by
trained and qualified technical service personnel, including functional safety checks; the
specific inspection items are as follows:
Environment and power meet the requirements.
Device and accessories have no mechanical damage.
The power supply has no wear, and the insulation is good.
Specified accessories are used.
Alarm system is functioning correctly.
Battery performance meets the requirements.
Monitoring functions are in good working condition.
Ground impedance and leakage current meet the requirements.