文件编号NR-TP-0715-008 版本1.0
- 85 -
SpO2 alarm preset
limits
Upper alarm limit
1%~100%
Lower alarm limit
0%~99%
SpO2 alarm preset
accuracy
±1%
SpO2 alerting signal
generates a delay
No delay
SpO2 value refresh
period
1s/time
SpO2 value refresh
delay
10s
Average period
Low sensitivity
78s
Intermediate sensitivity
46s
Advanced sensitivity
23s
Alarm condition delay
period
Low sensitivity
8s
Intermediate sensitivity
6s
Advanced sensitivity
3s
Alarm sign generates
delay period
0s
PR
Measuring range
30~254bpm
Resolution
1%
Accuracy
±2% or ±2bpm

Table of Contents