Chapter 4 Inspection
45
VISERA XENON LIGHT SOURCE CLV-S40
Operating the automatic brightness control
When this instrument is combined with any of the following products, be sure to
inspect the automatic brightness control.
This function can be inspected when the brightness control indicator is lit.
− OTV-S5
− OTV-S6
− OTV-SX
− OTV-SX2
− Video system that uses a BNC cable
− Combination of the OTV-S7V and a VISERA videoscope
(ENF TYPE V, CYF TYPE V, CYF TYPE VA or HYF TYPE
V)
Before disconnecting the video system from the endoscope,
be sure to press the standby switch to disable the light, or
minimize the brightness using manual brightness control. If
the video system is disconnected without performing one of
these operations, the light may cause operator eye injury or
burns.
• For further details, refer to the instruction manuals of the rigid
endoscope, fiberscope or videoscope and video system you
are using in combination with this unit.
• When switching the normal intensity mode to high intensity
mode, be sure to set the brightness level to below 0.
Otherwise, the brightness will exceed the necessary level,
and operator and/or patient injury may result.
• In combination with some non-Olympus video system, the
automatic brightness control may not operate properly, and
endoscopic observation and treatment may be impossible.