• When this instrument and the VISERA video system center
OTV-S7V are combined with any of the following products,
the brightness control indicator and MANUAL indicator are
disabled, and it is not possible to control the brightness from
this instrument. In this case, use the exposure level switch on
the OTV-S7V.
Refer to the instruction manual of the OTV-S7V.
− OTV-S7V camera head
− LTF TYPE V3
− A50001A/03A/21A/23A
• When using the automatic brightness control function with
the following products, brightness adjustment by the
brightness control switches are not available. In this case, set
the brightness level to “0”. For the exposure control, refer to
the instruction manual of the following products.
− OTV-S6
− OTV-SX2
1. Attach the video adapter to the endoscope’s eyepiece as shown in Figure
4.11 below (not necessary when using videoscopes).
Figure 4.11
2. Press the AUTO/MANUAL brightness selector and select the automatic
brightness control.