4. Setting the Unit (Unit Operation)
4-20
(31) Rate/CO
2
Conversion Value Setting (RATE)
Specifies the electricity rate or CO
2
emission level per 1kWh.
The range of selection (Numeric input):
00.000 to 99.999
Initial value: 0.000
Note
The rate/CO
2
conversion value setting is not possible when ETC is set to "OFF".
(32) Conversion Unit Setting (CONV)
Specifies the unit for the rate/CO
2
conversion setting (RATE).
The range of selection (Options):
JPY (Japanese yen) / USD (US dollar) / EUR (Euro) / CNY (Chinese yuan) / KRW
(Korean won) /
CO
2
(CO
2
emission level per 1kWh)
Initial value: JPY
Note
The conversion unit cannot be specified when ETC is set to "OFF".
(33) Power Startup REC Restoration (REREC)
Specifies if the REC restoration function is used in the event of power startup.
The range of selection (Options):
OFF / ON
Initial value: OFF
Setting Value
Operation
OFF
The unit does not automatically start recording at the next startup
after a power failure during recording.
ON
The unit automatically writes data to a memory card and starts
recording at the next startup after a power failure during recording.
Note
The power startup REC restoration setting is not possible when ETC is set to "OFF".
Important
An error occurs if an SD memory card is not inserted at the startup immediately after a
power failure when the REC restoration is set to "ON". The unit fails in writing data to the
memory card and does not start recording. To solve this situation, cancel the error state and
start recording via key operation.