6. Ratings and Performance
6-1
6. Ratings and Performance
Description
ZN-CTS
1-
A, ZN-CTM
1-
A
7-seg. 5-digit 2-step LCD display, auxiliary information indicator
displays
1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 1 min.
*1
Momentary power, Integrated power consumption
Normal mode, Sleep mode*
3
, High-speed logging mode
Continue mode*
4
, Ring mode*
5
Alarm output (Photocoupler output) *
6
Internal memory: approx. 6500 data items (The number of CHs is 1)
SD card with SDHC compatibility
*7
(Save measured values and
converted values; save and read setting values)
DC input: 24 VDC±10%
AC adapter: 100 to 240 VAC/50 to 60 Hz
Batteries: Two AAA batteries*
8
Max. 80 mA (AC adapter used)
Approx. 1 week*
10
Temperature
Battery Supply: -10
to 60
(no condensation or icing)
AC Adapter: 0
to 40
(no condensation or icing)
20 to 85%RH (no condensation or icing)
-15
to +60
, 20 to 85%RH (no condensation or icing)
20 MΩ (with DC500V Megger)
1000 VAC, 50/60 Hz, 1 min.: Between the case and I/O terminal
With mounting screws: 10 to 150 Hz, 0.7 mm double amplitude,
acceleration: 50 m/s
2
for each in X, Y and Z directions for 80 min.
With mounting magnets: 10 to 55 Hz,
0.3 mm double amplitude,
acceleration: 20 m/s
2
for each in X, Y and Z directions for 50 min.
150 m/s2 in 6 directions (+/-X, +/-Y, and +/-Z directions), 3 times
each*
11
ABS
IP30
Magnet mounting, screw mounting, hook, free standing
117.2 × 24.6 × 56.8 mm (Excluding protrusions)
Approx. 500 g
Instruction Sheet, Startup Guide, Mounting Magnets*
12
, Alarm Output
Connector*
13
, AC Adapter, DC cable, Ferrite core
*1: In high-speed logging mode, data is recorded in 83 ms at 60 Hz and in 100 ms at 50 Hz.
*2: Momentary power and integrated power values are converted from the measured current.
Correctly specify the number of used channels, applicable measurement target circuit, CT
type, frequency, voltage and power factor.
*3: The display turns OFF after 10 seconds of no user operation and recovers by a key
operation when SLEEP mode is specified. LAN cannot be used when sleep mode is
specified.
*4: Automatically writes the data to the SD memory card when the internal memory reaches its
capacity and continues recording until the SD card memory capacity reaches its limit. The
unit stops operation if there is no SD memory card inserted when the internal memory
reaches its capacity. (Recording can be resumed after inserting an SD memory card and

Table of Contents