Table of Contents

Questions and Answers

  • J
    Jasmine SanchezJul 29, 2025
    Why does the Omron Monitor display '-----' after I turn it on?
    • S
      Sara DavidsonJul 29, 2025
      The display may remain on “-----” after power on due to several reasons. First, the startup compensation timer setting might be too long. It can be set up to 99.9 s, so change the setting to an appropriate value. Second, the HOLD input might be ON. Turn OFF the HOLD input, as the display will remain on “-----” while it is active after power is turned on. Finally, check the RESET input and ensure it is turned OFF.
  • R
    Robert GentryAug 2, 2025
    What to do if the Omron K3HB-H input value is outside the measurement range?
    • I
      ispencerAug 2, 2025
      If the input value is outside the possible measurement range or the input is faulty, you can try the following. First, change the input type setting to an appropriate value in the initial setting level. Second, quickly return the input to within the possible measurement range. If the problem persists after these measures, repair may be necessary, so contact the point of purchase or your OMRON representative.
  • A
    Ashlee SmithAug 7, 2025
    What happens if the Omron Monitor measurement value is out of scale?
    • R
      Raymond PerezAug 7, 2025
      If the measurement value after scaling exceeds the range (greater than 99,999 or less than ?19,999), the operation will continue with a measurement value of 99,999 or ?19,999. If there is an operating problem, adjust the input range and scaling value until the measurement value falls within the range and place the switch below the E slot toward the front (K3HB-H only). Also, the scaling value may be inappropriate, so review it in the initial setting level.
  • A
    Aaron TerrellAug 11, 2025
    Why doesn't the Omron K3HB-H comparative output turn off?
    • P
      Phillip PenningtonAug 11, 2025
      If the comparative output of your Omron Monitor doesn't turn OFF even when the measurement value is back to normal, it might be due to the hysteresis setting being too large; change the setting to an appropriate value. Alternatively, check if the “output refresh stop” is set and turn it OFF.
  • T
    Timothy GomezAug 15, 2025
    What to do for non-volatile memory error on Omron Monitor?
    • I
      igreenAug 15, 2025
      Press the L [LEVEL] Key in this state for at least 3 s to return to the factory settings. If the problem still persists, repair is necessary. Contact the point of purchase or your OMRON representative.
  • K
    Kristen GordonAug 19, 2025
    What to do if Omron K3HB-H detects an unexpected Unit?
    • J
      John WilkinsonAug 20, 2025
      The mounting position depends on the Unit model. Check the Unit's model number and mount it in the correct position.
  • D
    Daniel CarterAug 23, 2025
    What to do if Omron K3HB-H displays an error after mounting a new unit?
    • K
      Kathryn WardAug 23, 2025
      If this is displayed the first time power is turned ON after mounting a new Unit, press the L [LEVEL] Key for at least 3 s to register the new Unit configuration.

Summary

Preface

Warranty and Limitations of Liability

Safety Precautions

Precautions for Safe Use

Section 1 Outline

1.1 Main Functions and Features of the K3 HB

Details the primary functions and features of the K3HB digital indicator.

Section 2 Preparations

2.1 Mounting

Covers external dimensions, panel cutout dimensions, and mounting methods.

2.2 Using I;O

Explains wiring diagrams for various outputs, inputs, and communication interfaces.

Section 3 Basic Application Methods

Section 4 Initial Setup

4.1 K3 HB-X Initial Setup Example (K3 HB-XVD)

Provides a step-by-step example for the initial setup of the K3HB-X.

4.2 K3 HB-V Initial Setup Example (K3 HB-VLC)

Provides a step-by-step example for the initial setup of the K3HB-V.

Section 5 Functions and Operations

Section 6 User Calibration

Section 7 Troubleshooting

7.1 Error Displays

Lists potential error displays and their descriptions.

7.2 Countermeasures

Provides troubleshooting steps for common symptoms.

Omron K3HB-H Specifications

General IconGeneral
Input TypeThermocouple, RTD, Voltage, Current
Number of Inputs1
Measurement RangeVaries by input type
Power Supply100 to 240 VAC
Output TypeRelay
CommunicationRS-485
Number of Outputs2
Operating Temperature-10 to 55 °C
Storage Temperature-25 to 65°C