Initialization
Section 4 Initial Setup
4-2
4.1 K3HB-X Initial Setup Example (K3HB-XVD)
The initial setup is explained in the following example.
Settings Example
In the following setting example, 1 to 5-V input is scaled to the range
0.000 to 1.000.
• If the measurement value goes above 0.700, comparative output
H turns ON.
• If the measurement value goes below 0.500, comparative output L
turns ON.
Initial Setup Flow
● Note ● To change a set value, press the S [SHIFT] Key. (The digit that can be
changed will flash.) Use the S [SHIFT] Key to move to the digit to be
changed, and change the setting using the U [UP] Key.
• If the display flashes “s.err” this indicates that the input is outside
the set range, and does not indicate product failure.
● Note ●
1. Move to the initial setting level by pressing the L [LEVEL] Key for at least
3 s (operation will stop).
2. Set input type A “in-ta” to “d vd” and press the M [MODE] Key twice.
1. Set scaling input value A1 “inp.
a1” to “1. 0000" and press the M [MODE]
Key.
2. Set scaling display value A1 “dsp.
a1” to “0” and press the M [MODE] Key.
3. Set scaling input value A2 “inp. a2” to “5. 0000" and press the M [MODE]
Key.
4. Set scaling display value A2 “dsp. a2” to “1000” and press the M [MODE]
Key.
1. Set the parameter “dp” to “,,.
,,,” and press the M [MODE] Key.
1.0000 V
5.0000 V
0.000
1.000
0.500
0.700
Display value
Input
value
Comparative output H
Comparative output L
Measurement
value
When the power is turned ON,
a number may be displayed that
is unrelated to the input range
setting.
To display correct values, the
correct input type must be
selected for the wiring.
A Check the wiring and turn the power ON.
Do not change the order of step
B.
When input type A is set, the
scaling value and decimal point
position will be initialized
automatically.
B Set input type A to 1.0000 to 5.0000 V.
C Set the scaling value.
D Set the decimal point position.