F
-
4300 CLAMP
-
ON ULTRASONIC FLOW METER
ONICON Incorporated 727.447.6140 Page 51 onicon.com
Log Site ID – Provide a number to the log site you’re creating. Can be any
number between 00-99.
Mode – 
Once option selected, press to store the setting. This setting cannot be

mode.
File Format – 


Excel spreadsheet for manipulation and graphing.
Date – 
Time –
Interval – 
Options: 10 sec, 30 sec, 1 min, 2 min, 5 min, 10 min, 15 min, 30 min, and 60 min.
At the fastest interval, 10 sec, the logger will last 8.5 years before it is full.


Although the logger storage capability is quite extensive, ONICON recommends

could present data transfer problems and corrupted data.
Data Log 
4.5 PROGRAMMING MENU PAGES (Continued)
--Data Logging-------
Log Site ID 01
Mode Flow
File Format .LG2
Date Mar. 20/2020
Time 12:28:41
Interval 30sec
Data Log Logging
IMPORTANT NOTE:
You MUST Delete an old log and Start a new log AFTER having made chang-
es to Log Site ID, Mode, Date, Time and/or Interval for those changes to be
applied.
IMPORTANT NOTE
Changing any of the parameters in the Units/Mode menu will start a new
log. It is recommended that you Delete and start a new log after changing
any Units/Mode settings.