9. Record the information shown
Record readings shown for: Flow rate and Flow total
10.
pulse outputs.
Current Output:
measured analog signals:
measured current in mA - 4
16
or,
measured voltage
5
record time interval between contact closures.)
11. Compare various output
signals to each other and to
control system.
rate indicated by the control system. Refer to the troubleshooting section of this
manual when readings are inconsistent.
End of standard commissioning.
7.3 COMMISSIONING WORKSHEET
readings as quickly as possible.
STEP TEST/MEASUREMENT S/N: S/N: S/N: S/N:
1.
2.
3.
4.
5. Control system
6.
7.
wiring Ok:
8. Record measured
supply voltage:
9. Record the
displayed data:
10. Analog or pulse
output(s)
4-20 mA signal:
11. Flow rate displayed
by control system.
7.2 COMMISSIONING PROCEDURE (Continued)