CAUTION
Cleaning materials
Wear powder-free gloves at all times when cleaning the device including the scan head
and main mirror. If the gloves become contaminated they must be replaced immediately.
Do not use lint cloths, tissues, or other materials that may create dust, on or near the scan
head. Always allow cleaned areas to air-dry.
Improper cleaning methods or use of inappropriate cleaning and disinfecting agents can
result in the following:
l Damaged equipment
l Poor imaging performance
l Increased risk of electric shock
Always follow the cleaning and disinfecting agents’ manufacturer’s instructions and
precautions; for storage, method of application, protective gloves, shelf life and disposal
to help assure effective and safe use of the product.
Be careful not to damage the scan head covers when cleaning.
Do not use harsh detergent, abrasive cleaners or solvents that are not recommended by
the manufacturer.
Do not expose equipment parts to steam or sterilizers as this will damage the device.
Do not allow liquids to get into the equipment.
If you believe cleaning sprays or liquids have got into the equipment, disconnect the
power supply and have the equipment checked by Optos qualified service personnel.
Do not let the cleaning wipes come into contact with the inside of the instrument.
Do not use the equipment until you have confirmed it is safe to use.
1.8 Maintaining information security
WARNING
Do not connect this equipment with any other equipment not provided in the accessory kit.
Do not use any cables, adapters, or products that are not provided or approved for the
device as this may result in non-compliance.