CAUTION
Wear powder-free gloves at all times when cleaning the device including the scan head
and main mirror. If the gloves become contaminated they must be replaced
immediately.
In order to prevent any contamination always clean and disinfect parts of the equipment
which will be in contact with the patient (see Cleaning before each patient on the facing
page).
Adequate cleaning and disinfection is necessary to prevent disease transmission.
Thoroughly clean and disinfect the equipment surfaces that contact the patient and all
equipment surfaces likely to become soiled during use.
Improper cleaning methods or use of inappropriate cleaning and disinfecting agents can
result in the following:
l Damaged equipment
l Poor imaging performance
l Increased risk of electric shock
Always follow the cleaning and disinfecting agents’ manufacturer’s instructions and
precautions; for storage, method of application, protective gloves, shelf life and disposal
to help assure effective and safe use of the product.
Do NOT use lint cloths, tissues, or other materials that may create dust, on or near the
scan head. Always allow cleaned areas to air-dry.
Do NOT use harsh detergent, abrasive cleaners or solvents that are NOT recommended
by the manufacturer.
Do NOT expose equipment parts to steam or sterilizers as this will damage the device.
Do NOT allow liquids to get into the equipment.
If you believe cleaning sprays or liquids have got into the equipment, disconnect the
power supply and have the equipment checked by Optos qualified service personnel.
Do NOT let the cleaning wipes come into contact with the inside of the instrument.
Do NOT use the equipment until you have confirmed it is safe to use.