Chapter 8 - Technical Specification
TABLE 15: Environmental ranges
Status Temperature
Relative Humidity
(Non-condensing)
Atmospheric Pressure
Normal use +10°C to +35°C 75% 70kPa to 106kPa
Storage -10°C to +50°C 95% 70kPa to 106kPa
Transport -10°C to +50°C 95% 50kPa to 106kPa
Note
Device is DEHP free.
9.2 Laser safety designation (IEC 60825-1)
CAUTION
In accordance with IEC 60825-1 Ed. 3, the Laser Management System maintains a
Class 1 accessible emission limit (AEL) at the eye for normal operation and under
foreseeable fault conditions. Based on current scientific knowledge, a Class 1 laser
device can be considered as safe by engineering design.
Wavelengths
SLO: 532 ±5 nm, 635 ±5 nm
SLO fa: 488 ±2 nm
SLO icg: 802 ±2 nm
Swept source OCT wavelength: 1050 ±10 nm
Resolution
Typical Axial Resolution: <7 micron
Transverse Resolution: 20 micron (in tissue)
optomap plus: 14 micron
9.2.1 External Laser warning information
Location of Laser warning Warning information
Positioned on the rear scan
head cover.
CLASS 1 LASER DEVICE
Part Number: G110230-003-ENG Page 73 of 84
Copyright 2019-2021, Optos plc. All rights reserved. English