Chapter 8 - Technical Specification
9.2.2 Internal Laser warning information
Location of Laser warning Warning information
Positioned on the laser
module (BDIO)
VISIBLE AND INVISIBLE LASER RADIATION
AVOID DIRECT EXPOSURE TO BEAM
CLASS 3B LASER PRODUCT
Positioned on the swept
source cover
WARNING - LASER RADIATION
AVOID EXPOSURE TO BEAM
CLASS 3B LASER PRODUCT
Positioned on the fiber
network cover
CAUTION - INVISIBLE LASER RADIATION
PRESENT IF FIBER REMOVED, CLASS 3R
LASER RADIATION
Positioned on the reference
arm cover
WARNING AVOID EXPOSURE TO BEAM,
INVISIBLE RADIATION PRESENT WHEN COVER IS REMOVED,
CLASS 3B LASER RADIATION
9.3 Electromagnetic compatibility (EMC)
The P200TxE is intended for use in the electromagnetic environment as specified in the following tables.
The customer or the user of the device should assure that it is used in such an environment.
9.3.1 Guidance and manufacturer declaration electromagnetic emissions
TABLE 16: List of electromagnetic emission compliance
Emissions test
IEC60601
compliance
test level
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The P200TxE uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are NOT likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The P200TxE is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Pst
Dmax
9.3.2 Guidance and manufacturer declaration electromagnetic immunity
TABLE 17: List of electromagnetic immunity compliance
Immunity test
IEC60601 compliance
test level
Electromagnetic environment-guidance
Electrostatic
discharge
IEC61000-4-2
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Page 74 of 84 Part Number: G110230-003-ENG
English Copyright 2019-2021, Optos plc. All rights reserved.