Error message Error no. Content Cause Possible measure
M1 Brake Error 1505 Brake error on left
motor
M2 Brake Error 1506 Brake error on right
motor
e.g.defective plug
connection, cable
break
Defective brake
Check plug connec
tions and cable to
brake
If error persists: con
tact specialist dealer
M1 Motor Error 3B00 Left motor not con
nected
M2 Motor Error 3C00 Right motor not con
nected
e.g.defective plug
connection, cable
break
Check plug connec
tions and cable to
motor
Inhibit Active 1E01; 1E20; 1E21;
1E22; 1E23
Driving function
blocked due to
external factors
Battery charger may
be connected
Disconnect battery
charger
Brake Lamp Short (None) Short circuit in brake
light electrical circuit
e.g.defective plug
connection, cable
break
Defective brake light
Check plug connec
tions and cable to
brake light
If error persists: con
tact specialist dealer
Left Lamp Short 7205 Short circuit in elec
trical circuit of left
front/rear light
Right Lamp Short 7209 Short circuit in elec
trical circuit of right
front/rear light
e.g.defective plug
connection, cable
break
Defective front/rear
light
Check plug connec
tions and cable to
front/rear light
If error persists: con
tact specialist dealer
L Ind Lamp Short 7206 Short circuit in elec
trical circuit of left
direction indicator
R Ind Lamp Short 720A Short circuit in elec
trical circuit of right
direction indicator
e.g.defective plug
connection, cable
break
Defective direction
indicator signal
Check plug connec
tions and cable to
direction indicator
If error persists: con
tact specialist dealer
L Ind Lamp Failed 7207 Error in electrical cir
cuit of left direction
indicator
R Ind Lamp Failed 7208 Error in electrical cir
cuit of right direction
indicator
e.g.defective plug
connection, cable
break
Defective bulb
Check plug connec
tions and cable to
direction indicator
If error persists: con
tact specialist dealer
Over-current (None) Amount of current in
an actuator circuit
too great
e.g.defective plug
connection, cable
break
Defective servomotor
(actuator) or control
ler
Check plug connec
tions and cable to
servomotor; verify
that actuating drive
can move freely
If error persists: con
tact specialist dealer
Overtemp. (Acts) (None) Temperature on actu
ator circuits on con
troller too high
e.g.defective plug
connections
Defective servomotor
(actuator)
Cool down phase
Check plug connec
tions and cable to
actuator
In case of recur
rence: contact spe
cialist dealer
Overtemp. (Lamps) (None) Temperature on the
light circuit on the
controller too high
e.g.defective plug
connections
Defective lamp
Cool down phase
Check plug connec
tions and cable to
light
In case of recur
rence: contact spe
cialist dealer
107Juvo
Maintenance and repair

Table of Contents