Error message Error no. Content Cause Possible measure
Memory Error (None) Unspecified memory
error in control mod
ules
e.g.defective plug
connection, cable
break
Defective control
module
Contact specialist
dealer
PM Memory Error (None) Memory error in con
troller
e.g.defective plug
connection, cable
break
Defective controller
Check all plug con
nections and cables
on controller
If error persists: con
tact specialist dealer
Bad Cable (None) Error on the commu
nication cables
between control
modules
e.g.defective plug
connection, cable
break
Check all plug con
nections and cables
on the control mod
ules (incl. controller)
If error persists: con
tact specialist dealer
Module Error (None) Error in control mod
ules
e.g.unspecified error
in displayed control
module
Displayed control
module defective
Check all plug con
nections and cables
If error persists: con
tact specialist dealer
System Error (None) System error Non-attributable error
in control modules
(primarily from third-
party providers)
Check all plug con
nections and cables;
disconnect/test any
modules from third-
party providers in
succession if applic
able
If error persists: con
tact specialist dealer
SID Detached (None) Special control dis
connected
Special control dis
connected from LCD
monitor (sip and puff
control)
Special control con
nection to LCD mon
itor defective
Check all plug con
nections and cables
between special con
trol connection and
sip and puff control;
check programming
parameters
If error persists: con
tact specialist dealer
1E07 External on/off switch
disconnected from
control panel
Switch Detached
1E08 External profile/mode
switch disconnected
from control panel
e.g.defective cable
connection to control
panel
Defective switch or
control panel
Check cable connec
tion and switching
function of the extern
al switch
If error persists: con
tact specialist dealer
Gone to Sleep (None) Control unit switches
to sleep mode
Extended inactivity
while control unit is
switched on
Switch control unit
off/on
Charging (None) Battery charger is
connected
Battery charger is
connected to char
ging receptacle
Disconnect battery
charger prior to driv
ing
JS Static Timeout (None) Joystick holding time
exceeded
Excessive joystick
deflection detected
(control unit stops
drive system to pre
vent potential dam
age to the motors)
Switch control unit
off/on
If error persists:
replace control panel
/ joystick; contact
specialist dealer
108
Maintenance and repair
Juvo