Error message Error no. Content Cause Possible measure
Memory Error (None) Unspecified memory
error in control mod
ules
e.g.defective plug
connection, cable
break
Defective control
module
Contact specialist
dealer
PM Memory Error (None) Memory error in con
troller
e.g.defective plug
connection, cable
break
Defective controller
Check all plug con
nections and cables
on controller
If error persists: con
tact specialist dealer
Bad Cable (None) Error on the commu
nication cables
between control
modules
e.g.defective plug
connection, cable
break
Check all plug con
nections and cables
on the control mod
ules (incl. controller)
If error persists: con
tact specialist dealer
Module Error (None) Error in control mod
ules
e.g.unspecified error
in displayed control
module
Displayed control
module defective
Check all plug con
nections and cables
If error persists: con
tact specialist dealer
System Error (None) System error Non-attributable error
in control modules
(primarily from third-
party providers)
Check all plug con
nections and cables;
disconnect/test any
modules from third-
party providers in
succession if applic
able
If error persists: con
tact specialist dealer
SID Detached (None) Special control dis
connected
Special control dis
connected from LCD
monitor (sip and puff
control)
Special control con
nection to LCD mon
itor defective
Check all plug con
nections and cables
between special con
trol connection and
sip and puff control;
check programming
parameters
If error persists: con
tact specialist dealer
1E07 External on/off switch
disconnected from
control panel
Switch Detached
1E08 External profile/mode
switch disconnected
from control panel
e.g.defective cable
connection to control
panel
Defective switch or
control panel
Check cable connec
tion and switching
function of the extern
al switch
If error persists: con
tact specialist dealer
Gone to Sleep (None) Control unit switches
to sleep mode
Extended inactivity
while control unit is
switched on
Switch control unit
off/on
Charging (None) Battery charger is
connected
Battery charger is
connected to char
ging receptacle
Disconnect battery
charger prior to driv
ing
JS Static Timeout (None) Joystick holding time
exceeded
Excessive joystick
deflection detected
(control unit stops
drive system to pre
vent potential dam
age to the motors)
Switch control unit
off/on
If error persists:
replace control panel
/ joystick; contact
specialist dealer