Device specific errors
Error
Number
Error Description description/explanation/examples
(1)220
Parser error Generic error in the parser.
(1)221
Illegal parser call The parser was called when it should not be active.
(1)222
Unrecognized input
character
A character not in the valid IEEE488.2 character set
was part of a command.
(1)223
Internal parser error The parser reached an unexpected internal state.
(1)230
Response formatter
error
Generic error in the response formatter.
(1)231
Bad response for
-
matter call
The response formatter was called when it should
not be active.
(1)232
Bad response for
-
matter call (eom)
The response formatter was called to output an end
of message, when it should not be active.
(1)233
Invalid function
code to response
formatter
The response formatter was requested to output
data for an unrecognized function.
(1)234
Invalid header type
to response format-
ter
The response formatter was called with bad data for
the response header (normally empty)
(1)235
Invalid data type to
response formatter
The response formatter was called with bad data for
the response data.
(1)240
Unrecognized error
number in error
queue
An error number was found in the error queue for
which no matching error information was found.
Error Messages
Error Code (1)220 to (1)240 7-15

Table of Contents

Other manuals for Pendulum CNT-90 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 Series and is the answer not in the manual?

Pendulum CNT-90 Series Specifications

General IconGeneral
BrandPendulum
ModelCNT-90 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals