Command Reference 8-63
:MEASure :FREQuency :RATio?
[8 [<expected value> [,<resolution>]][,<(@«1|2|3»)>,<(@«1|2|3»)>]]
Frequency Ratio
Frequency ratio measurements between two inputs.
Example:
SEND®
:MEAS:FREQ:RAT? 8 (@1),(@3)
READ¬ 2.345625764333E+000
This example measures the ratio between input A and input C.
+
The channel is expression data and must be within parentheses ( ).
Parameters: <expected value> and <resolution> are ignored
<(@«1|2|3»)>,<(@«1|2|3»)> are the measurement channels:
(@1) means input A
(@2) means input B
(@3) means input C (RF input option)
If you omit the channels, the instrument measures between input A and input B.
Complies with standards: SCPI 1991.0, confirmed.
:MEASure [:VOLT] :NCYCles?
[8 [(@1|@2|@3)]
Number of Cycles in Burst
If :FREQ:BURSt is active, this function measures the number of cycles in each
burst.
Returned format: <Numeric value (integer)>
Example:
SEND® :MEAS:NCYC? (@3)
READ¬
2356¿
This example shows a measurement on the RF channel.
+
The channel is expression data and must be within parentheses ( ).
<(@«1|2|3»)>,<(@«1|2|3»)> are the measurement channels:
(@1) means input A
(@2) means input B
(@3) means input C (RF input option)

Table of Contents

Other manuals for Pendulum CNT-90 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 Series and is the answer not in the manual?

Pendulum CNT-90 Series Specifications

General IconGeneral
BrandPendulum
ModelCNT-90 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals