Command Reference 8-139
*SAV
_<Decimal data>
Save
Saves the current settings of the instrument in an internal nonvolatile memory.
Nineteen memory locations are available. Switching the power off and on does not
change the settings stored in the registers.
Note that memory positions 1 to 10 can be protected from the front panel USER
OPT menu. If this has been done, use the :SYSTem:UNPRotect command to al
-
ter these memory positions.
Parameters
<Decimal data> = a number between 1 and 19.
Example:
SEND®
*SAV 8 11¿
Complies with standards: IEEE 488.2 1987