Command Reference 8-107
:STATus :OPERation :CONDition?
Read Operation Status Condition Register
Reads out the contents of the operation status condition register. This register re
-
flects the state of the measurement process. See table below.
Returned Format:
<Decimal data> = the sum (between 0 and 368) of all bits that are true. See table below:
Bit No. Weight Condition
11 2048 Computing statistics
10 1024 In limit
9 512 Using internal reference
8 256 Meas. stopped / Computing statistics (in compatibility mode)
6 64 Waiting for bus arming
5 32 Waiting for triggering and / or external arming
4 16 Measurement started
0 1 Calibrating
Complies with standards: SCPI 1991.0, confirmed.