8-10 Command Reference
:ARM :STOP :SOURce
8 «EXTernal1 | EXTernal2 | EXTernal4 | TIMer | IMMediate»
External Stop Arming Source
Selects STOP arming input or switches off the STOP arming function. The CNT-91
has also a programmable timer that is accessible in Totalize mode.
Parameters:
EXTernal1 is input A
EXTernal2 is input B
EXTernal4 is input E
TIMe is timed STOP in Totalize measurements (CNT-91 only). The time is set with
the command :ARM:STOP:TIM 8 <stop delay time>.
IMMediate is Stop arming OFF
Returned format: EXT1|EXT2|EXT4|TIM|IMM¿
Example:
SEND®
:ARM:STOP:SOUR 8 EXT4¿
*RST condition: IMM
Complies with standards: SCPI 1991.0, confirmed.
:ARM :STOP :TIMer CNT-91
8 <Numeric value> | MIN | MAX
Setting Gate Time in Timed Totalize Measurements
This command sets a delay between a pulse on the selected start arming input, i.e.
when totalizing starts, and the point of time when totalizing stops.
Range: 20 ns to 2 s, 10 ns resolution.
Parameters:
<Numeric value> is a number between 20
*
10
-9
and2s.
MIN gives 20
*
10
-9
s.
MAX gives 2 s
Returned format: <Numeric value>¿
Example:
SEND® :ARM:STOP:TIM 8 0.1¿
*RST condition: 0

Table of Contents

Other manuals for Pendulum CNT-90 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 Series and is the answer not in the manual?

Pendulum CNT-90 Series Specifications

General IconGeneral
BrandPendulum
ModelCNT-90 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals