Command Reference 8-117
:SYSTem :TEMPerature?
Read Temperature
This command returns the temperature in°C at the fan control sensor inside the in-
strument housing.
Returned format: <Numeric value>¿
Example:
SEND® :SYST:TEMP?
READ¬ 50
:SYSTem :TALKonlyCNT-91
8 ON
Enter Talk Only Mode
The main purpose is to transfer streaming data fast in monitoring systems without
predefined limits for time or number of samples. It is a non-reversible command,
i.e. you can only return to normal bus mode by sending IFC or pressing the CAN
-
CEL button on the front panel.
The Talk Only output buffer can hold one value. If a new measurement result is
ready for output before the previous one has been transferred, the new value is
rejected and the previous transfer is left undisturbed.
Consequently a pause during the reading will cause the first value read after the
pause to be the first measurement finished after the latest pre-pause value was
read. The second value read will be that of the most recently finished measure
-
ment. All values in between are lost. The same applies if there's a pause between
turning on Talk Only and starting to read values. So, a dummy read is recommend
-
able in many cases.
Prerequisites
DISPlay:ENABle should be OFF. FORMat should be REAL or PACKed.
ARM:COUNt and TRIGger:COUNt should both be one. INIT:CONTinuous
should be ON.
Smart Period/Frequency/Time Interval or any functions using voltage measurement
or timestamp can not be used with Talk Only.

Table of Contents

Other manuals for Pendulum CNT-90 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 Series and is the answer not in the manual?

Pendulum CNT-90 Series Specifications

General IconGeneral
BrandPendulum
ModelCNT-90 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals