Command Reference 8-9
:ARM :SOURce
8 «EXTernal1 | EXTernal2 | EXTernal4 | IMMediate»
External Start Arming Source
Selects START arming input or switches off the start arming function. When
switched off the DELay is inactive.
Parameters:
EXTernal1 is input A
EXTernal2 is input B
EXTernal4 is input E
IMMediate is Start arming OFF
Note: For the Totalize function in the CNT-91, IMM means manual start-stop using
the commands :TOT:GATE 8 ON|OFF¿
Returned format: EXT1 | EXT2 | EXT4 | IMM¿
Example:
SEND®
:ARM:SOUR 8 EXT4¿
*RST condition: IMM
Complies with standards: SCPI 1991.0, confirmed.
:ARM :STOP :SLOPe
8 «POSitive | NEGative»
External Stop Arming Slope
Sets the slope for the stop arming condition.
Returned format: POS|NEG¿
Example:
SEND®
:ARM:STOP:SLOP 8 NEG¿
*RST condition: POS
Complies with standards: SCPI 1991.0, confirmed.