Command Reference 8-47
:INPut«[1]|2» :FILTer
8 <Boolean>
Analog Low Pass Filter
Switches on or off the analog low pass filter on input 1 (A) and/or input 2 (B). It has
a cutoff frequency of 100 kHz.
Parameters:
<Boolean> is «1 | ON» | «0 | OFF»
Returned format: 1|0¿
*RST condition OFF
Complies with standards: SCPI 1991.0, confirmed.
:INPut«[1]|2» :FILTer :DIGital
8 <Boolean>
Digital Low Pass Filter
Switches on or off the digital low pass filter on input 1 (A) and/or input 2 (B). The
cutoff frequency is set by the command:
:INPut«[1]|2»:FILTer:DIGital:FREQuency8<Numeric value>
Parameters:
<Boolean> is «1 | ON» | «0 | OFF»
Returned format: 1|0¿
*RST condition: OFF