Command Reference 8-95
:FREQuency :BURSt :SYNC :PERiod
8 «<Numeric value>|MIN|MAX»
Burst Sync Delay
Sets the synchronization delay time used in burst measurements. A correct value
should be longer than the burst time and shorter than 1/PRF, i.e. the inverse of the
pulse repetition frequency.
Parameters: <Numeric value> = a number between 1
*
10
-6
and2s.
Returned format: <Numeric value>¿
*RST condition: 400 ms
:FREQuency :BURSt :STARt :DELay
8 «<Numeric value>|MIN|MAX»
Burst Start Delay
Sets the burst start delay, i.e. the time length between the burst start and the actual
start of the burst measuring time. This parameter is used for controlling the point of
time when a measurement sample is taken.
Parameters: <Numeric value> = a number between 20 ns and 2 s.
Returned format: <Numeric value>¿
*RST condition: 200 ms