Command Reference 8-57
:MEASure :ARRay :<Measuring Function>?
8 (<array size>)[,[<parameters>] [,(<channels>)]]
Make an array of measurements
The :MEASure:ARRay query differs from the :MEASure query in that it performs
the number of measurements you decide in the <array size> and sends all the
measuring results in one string to the controller.
+
The array size for :MEASure and :CONFigure, and the channels, are expression
data that must be in parentheses ( ).
The :MEASure:ARRay query is a compound query identical to:
:ABORt;:CONFigure:ARRay:<Meas-func> 8 (<array-size>); :READ:ARRay? 8
(<array-size>)
Parameters:
<array size> sets the number of measurements in the array. The maximum number is limited to
10000 due to the physical size of the output buffer. See also FETCH:ARR? and
READ:ARR?
Returned format:
<Measuring result>{[,<measuring result>]}¿
Example:
SEND®
:MEAS:ARR:FREQ? 8 (10)
Ten measuring results will be returned.
Type of command:
Aborts all previous measurement commands if not *WAI is used, see page 8-142 .
Complies with standards: SCPI 1991.0, confirmed.

Table of Contents

Other manuals for Pendulum CNT-90 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 Series and is the answer not in the manual?

Pendulum CNT-90 Series Specifications

General IconGeneral
BrandPendulum
ModelCNT-90 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals