8-110 Command Reference
:STATus :QUEStionable :CONDition?
Read Questionable Data/Signal Condition Register
Reads out the contents of the status questionable condition register.
Returned Format:
<dec.data> = the sum (between 0 and 17920) of all bits that are true. See table below:
Bit No. Weight Condition
14 16384 Unexpected parameter
11 2048 Out of limit
10 1024 Measurement timeout / Out of limit (in compatibility mode)
9 512 Overflow
8 256 Calibration error
6 64 Phase interpolation calibration off
5 32 Frequency interpolation calibration off
2 4 Time interpolation calibration off
Complies with standards: SCPI 1991.0, confirmed.