8-48 Command Reference
:INPut«[1]|2» :FILTer :DIGital :FREQuency
8<Numeric value>|MIN|MAX
Set the Digital Low Pass Filter Cutoff Frequency
Any frequency between 1 Hz and 50 MHz can be entered. The filter is activated by
the command:
:INPut«[1]|2»:FILTer:DIGital8ON|OFF
Parameters:
<Numeric value> is a value between 1 and 50
*
10
6
MIN means 1 Hz
MAX means 50 MHz
Returned format: <Numeric value>¿
*RST condition: 100 kHz
:INPut«[1]|2» :IMPedance
8 «<Decimal data>|MAX|MIN»
Input Impedance
The impedance can be set to 50 W or 1 MW.
Parameters
MIN or <Decimal data> that rounds off to 50 or less, sets the input impedance to 50 W
MAX or <Decimal data> that rounds off to 1001 or more, sets the impedance to 1 MW.
Returned format:
5.00000000000E+001|1.00000000000E+6¿
Example for Input A (1)
SEND®
:INP:IMP 8 50
Sets the input A impedance to 50 W.
Example for Input B (2)
SEND®
:INP2:IMP 8 50
Sets the input B impedance to 50 W.
*RST condition 1MW
Complies with standards: SCPI 1991.0, confirmed.

Table of Contents

Other manuals for Pendulum CNT-90 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 Series and is the answer not in the manual?

Pendulum CNT-90 Series Specifications

General IconGeneral
BrandPendulum
ModelCNT-90 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals