8-92 Command Reference
:ACQuisition :APERture
8 «<Decimal value > | MIN | MAX»
Set the Measurement Time
Sets the gate time for one measurement.
Parameters: <decimal value> is 20 ns to 1000 s.
MIN gives 20 ns and MAX gives 1000 s.
Returned format:
<Decimal value >¿
*RST condition: 10 ms
SYST:PRESet condition: 200 ms
:ACQuisition :HOFF
8 <boolean>
Hold Off On/Off
Switches the Hold Off function On/Off.
Parameters:
<Boolean> = 1 / ON | 0 / OFF
Returned format: 1|0¿
*RST condition: OFF