8-4 Command Reference
:ABORt
Abort Measurement
The ABORt command terminates a measurement. The trigger subsystem state is
set to “idle-state”. The command does not invalidate already finished results when
breaking an array measurement. This means that you can fetch a partial result af
-
ter an abort.
Type of command:
Aborts all previous measurements if *WAI is not used.
Complies with standards: SCPI 1991.0, confirmed.