8-46 Command Reference
:INPut«[1]|2» :ATTenuation
8 «<Numeric value>|MAX|MIN»
Attenuation
Attenuates the input signal by 1 or 10. The attenuation is automatically set if the in
-
put level is set to AUTO.
Parameters:
<Numeric values> £ 5, and MIN gives attenuation 1.
<Numeric values> > 5, and MAX gives attenuation 10.
Returned format:
1.00000000000E+000|1.00000000000E+001 ¿
Example for Input A (1)
SEND®
:INP:ATT 8 10
Example for Input B (2)
SEND®
:INP2:ATT 8 10
*RST condition Input A (1) and Input B (2): 1 (but set by autotrigger since AUTO is on af
-
ter *RST.(:INP:LEV:AUTO 8 ON).
Complies with standards: SCPI 1991.0, confirmed.
:INPut«[1]|2» :COUPling
8 «AC|DC»
AC/DC Coupling
Selects AC coupling (normally used for frequency measurements), or DC cou-
pling (normally used for time measurements).
Returned format:
AC|DC¿
Example for Input A (1)
SEND®
:INP:COUP 8 DC
Example for Input B (2)
SEND®
:INP2:COUP 8 AC
*RST condition
Input A (1): AC
Input B (2): AC
Complies with standards: SCPI 1991.0, confirmed.

Table of Contents

Other manuals for Pendulum CNT-90 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 Series and is the answer not in the manual?

Pendulum CNT-90 Series Specifications

General IconGeneral
BrandPendulum
ModelCNT-90 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals