Command Reference 8-116
:SYSTem :SET
8 <Block data>
Read or Send Settings
Transmits in binary form the complete current state of the instrument. This data
can be sent to the instrument to later restore this setting. This command has the
same function as the *LRN? common command with the exception that it returns
the data only as response to :SYST:SET?. The query form of this command re
-
turns a definite block data element.
Parameters:
<Block data> is the instrument setting previously retrieved via the :SYSTem:SET? query.
Returned format: <Block data>¿
Example:
SEND® :SYST:SET?
READ¬ #41686<data byte 1><data byte 2>...<data byte 1686>
Note: The real number of data bytes will probably differ from the one specified
above and depends on the counter type and the firmware version.
Complies with standards: SCPI 1991.0, confirmed.
:SYSTem :PRESet
Preset
This command recalls the same default settings that are entered when you press
USER OPT ® Save/Recall ® Recall Setup ® Default.
+
These are not exactly the same settings as after
*
RST:
:SYST:PRES gives 200 ms Measurement Time
:SYST:PRES activates :INIT:CONT ON
*RST gives 10 ms Measurement Time
*RST activates :INIT:CONT OFF
See page 8-44.
See also: Default settings after *RST on page 2-2.