8-26 Command Reference
:CALibration :INTerpolator :AUTO
8 <Boolean>
Calibration of Interpolator
The '9X' is a reciprocal counter that uses an interpolating technique to increase the
resolution. In time measurements, for example, interpolation increases the resolu
-
tion from 10 ns to 0.1 ns.
The counter calibrates the interpolators automatically once for every measurement
when this command is ON. When this command is OFF, the counter does no cali
-
brations but uses the values from the last preceding calibration. The intention of
this command is to turn off the auto calibration for applications that dump measure
-
ments into the internal memory. This will increase the measurement speed.
Parameters
<Boolean> = ( 1 | ON / 0 | OFF )
Returned format: 1|0¿
*RST condition: ON