Specifications
Inputs A and B
Alternative data within parentheses refer to
input attenuator setting x10.
Frequency Range
DC-coupled:
1 M /25 pF or 50
(VSWR
2:1)
Max. Channel
Timing
Difference:
Residual Hyster-
esis after Com-
pensation:
30 mVp-p to 10 Vp-p
within ±5 V window
±15 (150) m V ± 1% of trigger
level
Automatically set to 50% of
input signal (10% and
90% for rise/fall time).
Relative level (in %) manually
adjustable when necessary.
Auto Hysteresis
Time Meas. :
Minimum hysteresis window
(+compensation)
70 % and 30 % of input
signal. Minimum hysteresis
window if arming on A or B is
activated.
Analog Noise
Reduction Filter:
1 Hz - 50 MHz using trigger
hold-off
350 V (DC+ AC
pk
) @ D C
to 440 Hz, falling to 12 VRMS
(x1) and 120 VRMS (x10) @ 1
MHz
Operating Input
Voltage Range
USER MANUAL ● CNT 9x Series ● Rev.22 February 2020