Using the Controls
Arm
start. A typical use is to clean up signals gen-
erated by bouncing relay contacts.
Fig. 2-14
CNT-90 & CNT-90XL: Setting arming
conditions.
Fig. 2-15
CNT-91(R): Setting arming conditions.
Arming is the general term used for the means
to control the actual start/stop of a
measurement. The normal free-running mode is
inhibited and triggering takes place when
certain pretrigger conditions are fulfilled.
The signal or signals used for initiating the
arming can be applied to three channels (A, B,
E), and the start channel can be different from
the stop channel. All conditions can be set via
this menu.
NOTE: Stop Delay can only be used for realizing the
function Timed Totalize in the CNT-91(R).
Trigger Hold-Off
Fig. 2-16
The trigger hold-off submenu.
A value input menu is opened where you can
set the delay during which the stop trigger
conditions are ignored after the measurement
2-12
USER MANUAL CNT 9x Series ● Rev.22 February 2020

Table of Contents

Other manuals for Pendulum CNT-90 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 Series and is the answer not in the manual?

Pendulum CNT-90 Series Specifications

General IconGeneral
BrandPendulum
ModelCNT-90 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals