Transducer Care Transducer Care and Operator Safety
128 Lumify Ultrasound System
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
CAUTION
Do not use paper products or products that are abrasive when cleaning the transducer. They
damage the soft lens of the acoustic window of the transducer.
CAUTION
During cleaning, disinfection, and sterilization, orient the USB connector and cable that must
remain dry higher than the wet parts, until all parts are dry. This helps keep liquid from
entering unsealed areas of the transducer.
CAUTION
When cleaning and disinfecting transducers, do not allow any fluid to enter electrical
connections or metal portions of the USB connector. Damage due to fluids in these areas is
not covered by the warranty or your service contract.
CAUTION
To keep fluids from entering the transducer, do not disconnect the USB cable from the
transducer during cleaning and disinfection.
CAUTION
When using an enzymatic cleaner, be sure to use the proper concentration and rinse
thoroughly.