Transducer Care and Operator Safety Transducer Care
Lumify Ultrasound System 129
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
CAUTION
Before storing transducers, ensure that they are thoroughly dry. If it is necessary to dry the
transducer lens or acoustic window after cleaning, use a soft cloth and a blotting motion,
instead of a wiping motion.
CAUTION
The only parts of the transducer that may be cleaned with isopropyl alcohol are the
transducer housing and lens or acoustic window. Ensure that the solution is only 70%
alcohol or less. Do not wipe any other part of a transducer with isopropyl alcohol (including
cables, USB connectors, or strain reliefs), as it can damage those parts of the transducer.
This damage is not covered by the warranty or your service contract.
Alcohol-Compatible Transducer Parts
The only part that can be wiped with 70% isopropyl alcohol or other alcohol-based disinfectant is the transducer
housing and lens outlined in the figure. Do not allow fluid to enter any unsealed area of the transducer.