Electromagnetic Compatibility Safety
Lumify Ultrasound System 59
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
NOTE
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Electromagnetic Immunity: Environment Guidance
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic
Environment Guidance
Electrostatic discharge
(ESD), IEC 61000-4-2
8 kV air discharge, 6 kV
contact discharge
8 kV air discharge, 6 kV
contact discharge
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast transient/
burst, IEC 61000-4-4
Not applicable. The device
does not function on AC
power
- - Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge, IEC 61000-4-5 Not applicable. The device
does not function on AC
power
- - Mains power quality should
be that of a typical
commercial or hospital
environment.