Safety Electromagnetic Compatibility
60 Lumify Ultrasound System
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic
Environment Guidance
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines, IEC 61000-4-11
Not applicable. The device
does not function on AC
power
- - Mains power quality should
be that of a typical
commercial or hospital
environment. If you require
continued operation during
power mains interruptions,
Philips recommends that
the system be powered
from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz) magnetic field,
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Conducted RF,
IEC 61000-4-6
3 VRMS 3 VRMS For recommended
separation distances, see
“Recommended Separation
Distance” on page 62.
Radiated RF, IEC 61000-4-3 3 V/m 3 V/m For recommended
separation distances, see
“Recommended Separation
Distance” on page 62.
Although most remote devices comply with their applicable standards for immunity, those
device requirements may not be as stringent as those required for medical equipment. It is the
responsibility of the installer and the user of this remote customer-supplied equipment to
ensure that it functions properly in the electromagnetic environment where the system is