CALCulate:AVERage:MINimum?
Read the minimum value found during the Min/Max operation. The multimeter
clears the value when Min/Max is turned on, when the power has been off or a
remote interface reset.
CALCulate:AVERage:MAXimum?
Read the maximum value found during the Min/Max operation. The multimeter
clears the value when Min/Max is turned on, when the power has been off or a
remote interface reset.
CALCulate:AVERage:AVERage?
Read the average value of all readings taken since the Min/Max operation has
been enabled. The multimeter clears the value when Min/Max is turned on,
when the power has been off or a remote interface reset.
CALCulate:AVERage:COUNt?
Read the number of readings taken since Min/Max has been enabled. The
multimeter clears the value when Min/Max is turned on, when the power has
been off or a remote interface reset.
CALCulate:NULL:OFFSet {<value>|MINimum|MAXimum}
Store a null value in the multimeter’s Null Register. You must turn on the math
operation before writing to the math register. You can set the null value to any
number between 0 and
±
120% of the highest range.
CALCulate:NULL:OFFSet?
Query the null value.
CALCulate:LIMit:LOWer {<value>|MINimum|MAXimum}
Set the lower limit for limit testing. You can set the value to any number from 0
to
±
120% of the highest range, for the present function.
CALCulate:LIMit:LOWer?
Query the lower limit for the limit testing.
CALCulate:LIMit:UPPer {<value>|MINimum|MAXimum}
Set the upper limit for limit testing. You can set the value to any number from 0
to
±
120% of the highest range, for the present function.