CALCulate:LIMit:UPPer?
Query the upper limit for the limit testing.
CALCulate:MXB:MMFactor {<value>|MINimum|MAXimum}
Set the value of M.
CALCulate:MXB:MMFactor? [MINimum|MAXimum]
Query the value of M.
CALCulate:MXB:MBFactor {<value>|MINimum|MAXimum}
Set the value of B.
CALCulate:MXB:MBFactor? [MINimum|MAXimum]
Query the value of B.
CALCulate:DB:REFerence {<value>|MINimum|MAXimum}
Store a relative value in the dB Relative Register. You must turn on the math
operation before writing to the math register. You can set the relative value to
any number between 0 dBm and
±
200 dBm.
CALCulate:DB:REFerence? [MINimum|MAXimum]
Query the dB relative value.
CALCulate:DBM:REFerence {<value>|MINimum|MAXimum}
Set the dBm reference value. Choose from: 50 ~ 8000 ohms.
CALCulate:DBM:REFerence? [MINimum|MAXimium]
Query the dBm reference value.
DATA:FEED RDG_STORE,{“CALCulate”|””}
Selects whether readings taken using the INITiate command are stored in the
multimeter’s internal memory (default) or not stored at all. In the default state
(DATA:FEED RDG_STORE,“CALC”), up to 2000 readings are stored in memory
when INITiate is executed. The MEASure? And CONFigure commands
automatically select “CALC”. With memory disabled (DATA:FEED
RDG_STORE,“”), readings taken using INITiate are not stored. This may be