You execute a *ESE 0 command.
※
Note: When the DMM is powered on, the standard event enable
register won’t be cleared if you have set the DMM previously by using
*PSC 0.
About the Questionable Data Register
The questionable data register reports the quality of the DMM’s
measurement results, as well overload conditions and high/low limit test
results. Through the enable register, all of the conditions can be reported
in the questionable data summary bit. At the moment, you have to write
a decimal value by using the STATus:QUESTionable:ENABle command to
set the enable register mask.
Bit Definitions for Questionable Data Register
Bit
Decimal
Value
Definition
0.
Voltage Overload
1
Range overload on dc/ac volts, frequency, period, diode or ratio
function
1.
Current Overload
2
Range overload on dc/ac current function.
2.
Not Used