EasyManua.ls Logo

Planmeca ProOne - Additional Safety Notes and Laser Radiation Information

Planmeca ProOne
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 SAFETY PRECAUTIONS
6 Planmeca ProOne User’s Manual
CAUTION
Never use a defective or damaged X-ray unit. Contact
your service technician for help.
CAUTION
Do not modify the X-ray unit. The X-ray unit must be
serviced by qualified personnel only.
NOTE
Before taking an exposure, ask any female patient of
childbearing age whether she might be pregnant. The X-
ray unit is not intended for use on pregnant women.
NOTE
If the X-ray unit has been stored at temperatures below
+10°C for more than a few hours, time must be allowed for
the unit to reach room temperature before turning it on.
NOTE
Ensure efficient air conditioning in the X-ray room. It is
recommended to keep the room temperature between
+20°C and +25°C at all times.
NOTE
If the X-ray unit is not connected to an Uninterruptible
Power Supply (UPS), switch the X-ray unit off during
lightning storms.
NOTE
FOR US USERS:
The patient positioning lights are class II laser products (21
CFR § 1040.10).
NOTE
FOR EUROPEAN USERS:
The patient positioning lights are class 1 laser products
(Standard IEC / EN 60825-1: 2007).
NOTE
EMC requirements have to be considered, and the
equipment must be installed and put into service according
to the specific EMC information provided in the
accompanying documents.
NOTE
Portable and mobile RF communications equipment can
affect the X-ray unit.
CAUTION
LASER RADIATION -
DO NOT STARE
INTO BEAM
1mW
635nm
CLASS II
LASER
PRODUCT
LBL-X-099
LBL-X-100
AVOID
EXPOSURE-
Laser radiation
is emitted from
this aperture.
CLASS 1 LASER PRODUCT
APPAREIL À LASER DE CLASSE 1
IEC 60825-1:2007

Table of Contents

Other manuals for Planmeca ProOne

Related product manuals