15 HELP MESSAGES
User’s Manual Planmeca ProOne 101
H06-003 Dimax sensor Radiation level The radiation level on Dimax sensor is
below average or factory settings.
If necessary, contact your service
technician to improve image quality.
H06-004 Beam The X-ray beam is not symmetrical.
If necessary, contact your service
technician to improve image quality. The
collimator needs to be readjusted.
H06-008 Status The Dimax sensor is not ready for
exposure.
Wait for a moment.
H06-009 X-ray tube Heat capacity The X-ray tube will overheat with the
selected exposure values (kV/mA).
Use lower exposure values or wait for
the X-ray tube to cool down.
H06-010 Tube head Temperature The temperature of the tube head is too
high.
Wait for a few minutes for the tube head
to cool down.
H06-014 DEC Segmenting The lower horizontal jaw segment
(partial exposure) cannot be selected if
DEC (Dynamic Exposure Control) is
used.
H06-015 DEC DEC not available DEC (Dynamic Exposure Control)
cannot be used if only the lower
horizontal jaw segment is selected
(partial exposure).
H07-002 Licences Code The entered licence code was wrong.
Check that you have got the correct
licence code for the desired feature on
this X-ray unit.
H07-003 DEC Calibration DEC (Dynamic Exposure Control) is not
available.
Contact your service technician for help.
DEC has not been calibrated.
H07-004 Target value DEC (Dynamic Exposure Control)
calibration value is out of limits.
Target value has to fall between 200
and 2000.
H08-001 Software
Software downgrade is not allowed.
CPU PCB version I2 or newer installed,
it is not possible to downgrade the
software from version 1.9.1.
Code Explanation Comments