Push and hold the level arm to its down most position. Adjust the tension of the gas spring by
rotating the adjustment screw with a 6 mm Allen key as shown on the figure below. Turning
the screw clockwise will increase the tension, i.e, the weight-carrying capacity.
Add the lifting friction of the level arm slightly so that the arm moves smoothly.