Planmeca Sovereign 5
PRE-INSTALLATION REQUIREMENTS
Installation manual
Unit with 3- or 5-function pylon with rotating joint
3.2 Basement
NOTE The dealer must check the floor material and strength of the installation site before in-
stallation.
The floor must be straight within 1%.
The concrete or stone floor must be 50 mm (2 in.) thick. The concrete strength must be type
C20/25-C50/60 (According standard EN 206-1:2000), non-cracked concrete. Otherwise the
installation plate must be used.
Antistatic floor materials should be preferred.
3.3 Environmental requirements
Transportation conditions
Accepted limits for unit transportation are the following:
Temperatures: -20 °C to +60 °C (-4 °F to 140 °F)
Relative humidity: 5% RH to 95% RH “Absence of condensation”
Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
REFERENCE JOINT
for balanced arm.
R1600
(63.0”)
R600
(23.6”)
Inner and outer joint
of balanced arm.
S261.eps