INSTALLING THE PYLON AND ARMS
54 Planmeca Sovereign
Installation manual
The OP delivery arm is attached to the pylon with eight attachment screws and washers (four
on both sides). Attach the screws to the both sides using a 6 mm Allen key.
After attaching the screws tighten them all.
Remove the OP delivery arm transport plate and the OP delivery arm middle joint transport
plate as described in section 8.3 “Installing the fixed OP delivery arm” on page 65.